Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

344

 For information on setting up the status outputs (see Section 3.3.1.4).

CONTROL INPUTS: By applying 5V DC power to these digital inputs from an external source such as
a PLC or Data logger zero point and span point calibrations can be remotely initiated. .

 For information on setting up the status inputs (see Section 3.3.1.6).

13.3.3.6. Internal Digital I/O

There are several internal digital control signals that are generated by the motherboard under CPU
control and used to control subsystems of the analyzer.

ELECTRICAL TEST CONTROL: When the CPU sets this control signal to high (ON) the electric
test feature (ETEST) is initiated (see Section 8.3).

 The

ETEST can be initiated by following the procedure in Section 12.7.12.2, or by setting the

SIGNAL I/O Function ELEC_TEST to ON.

OPTICAL TEST (OTEST) CONTROL: When the CPU sets this control signal to high (ON) the
optical test feature is initiated (see Section 8.3).

 The

OTEST can be initiated by following the procedure in 12.7.12.1, or by setting the SIGNAL I/O

Function OPTIC_TEST to ON.

PMT PREAMPLIFIER RANGE CONTROL: The CPU uses this control switch the instrument
between its LOW and HIGH physical ranges (see Section 5.4.1).

 The instrument can be forced into its HIGH physical range setting the SIGNAL I/O function

PREAMP_RANGE_HI to ON.

O

3

GEN STATUS: The CPU uses this control signal to turn the O

3

generator ON/OFF by setting it to

HIGH/LOW respectively. The CPU turns OFF the O

3

generator if there is if there is no or low air flow

to it as measured by the O

3

flow sensor or if the instrument has been turned off for more than 30 minutes.

 The

O

3

generator can be manually turned ON/OFF by using the OZONE GENERATOR OVERIDE

feature (See Section 12.7.15.1) or by setting the SIGNAL I/O function O3GEN_STATUS to ON or
OFF.

Note

Any I/O signals changed while in the signal I/O menu will remain in effect
ONLY until signal I/O menu is exited.

The analyzer regains control of these signals upon exit and returns them
to their normal value/setting.

13.3.3.7. I

2

C Data Bus

I

2

C is a two-way, clocked, bi-directional digital serial I/O bus that is used widely in commercial and

consumer electronic systems. A transceiver on the Motherboard converts data and control signals from
the PC-104 bus to I

2

C format. The data is then fed to the relay board, optional analog input board and

valve driver board circuitry.