Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

71

HNO

3

is a liquid at room temperature, so once the HNO

3

is released by the permeation tube it condenses

and collects along the 9110T’s wetted surfaces, While liquid HNO

3

does not directly effect the quality of

NO

x

measurements of the Model 9110T, it does give off small amounts of gaseous HNO

3

which is

converted into NO by the 9110T’s NO

x

 NO converter resulting in an artificially high NO

2

concentration by 8% to 12%. This is particularly bothersome when 9110T is attempting to measure a
zero point, such as during calibration, since the NO

2

concentration will only reach a true zero point once

the majority of the HNO

3

coating the wetted surfaces has reverted to NO

2

and this can take a very long

time.

The 9110T includes a special HNO

3

desorbed which eliminates any HNO

3

given off by the permeation

tube before it can be converted into NO by the analyzer’s converter.

PNEUMATIC LAYOUT FOR ZERO SCRUBBER AND IZS (OPT 50G)

Figure 3-27:

Pneumatics with the Internal Span Gas Generator (OPT 50G)

Table 3-11: Internal Span Gas Generator Valve Operating States OPT 50G

MODE VALVE

CONDITION

VALVE PORT

STATUS

SAMPLE

Sample/Cal

Open to SAMPLE inlet

3 2