C bus, Lcd/display module, Relay pca – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 285: Motherboard

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

285

12.7.3. I

2

C BUS

Operation of the I

2

C bus can be verified by observing the behavior of D1 on the relay PCA & D2 on the

Valve Driver PCA. Assuming that the DC power supplies are operating properly, the I

2

C bus is operating

properly if D1 on the relay PCA and D2 of the Valve Driver PCA are flashing

There is a problem with the I

2

C bus if both D1 on the relay PCA and D2 of the Valve Driver PCA are

ON/OFF constantly.

12.7.4. LCD/DISPLAY MODULE

TOUCHSCREEN INTERFACE

Assuming that there are no wiring problems and that the DC power supplies are operating properly, the
display screen should light and show the splash screen and other indications of its state as the CPU goes
through its initialization process.

12.7.5. RELAY PCA

The Relay PCA can be most easily checked by observing the condition of the status LEDs on the Relay
PCA (see Section 12.3.2), and using the SIGNAL I/O submenu under the DIAG menu (see Section
12.1.3) to toggle each LED ON or OFF.

If D1 on the Relay PCA is flashing and the status indicator for the output in question (Heater power,
Valve Drive, etc.) toggles properly using the Signal I/O function, then the associated control device on
the Relay PCA is bad.

Several of the control devices are in sockets and can be easily replaced. The following table lists the
control device associated with a particular function:

Table 12-8:

Relay PCA Control Devices

FUNCTION CONTROL

DEVICE

SOCKETED

All valves

U5

Yes

Reaction Cell Heater

K1

Yes

NO

2

 NO Converter heater

K2

Yes

Permeation Tube Heater for

Optional Internal Span Gas Generator

K4 Yes

12.7.6. MOTHERBOARD

12.7.6.1. Test Channel / Analog Outputs Voltage

The ANALOG OUTPUT submenu, located under the SETUP MORE DIAG menu is used to
verify that the 9110T analyzer’s three analog outputs are working properly. The test generates a signal
on all three outputs simultaneously as shown in the following table: