Password security for serial remote communications – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 191

Advertising
background image

Model 9110TH NOx Analyzer

Remote Operation

Teledyne Analytical Instruments

191

8.4. PASSWORD SECURITY FOR SERIAL REMOTE

COMMUNICATIONS

In order to provide security for remote access of the 9110T, a LOGON feature can be enabled to require
a password before the instrument will accept commands. This is done by turning on the SECURITY
MODE
(refer to Section 5.5). Once the SECURITY MODE is enabled, the following items apply.

 A password is required before the port will respond or pass on commands.

 If the port is inactive for one hour, it will automatically logoff, which can also be achieved with the

LOGOFF command.

 Three unsuccessful attempts to log on with an incorrect password will cause subsequent logins to

be disabled for 1 hour, even if the correct password is used.

 If not logged on, the only active command is the '?' request for the help screen.

 The following messages will be returned at logon:

 LOGON SUCCESSFUL - Correct password given

 LOGON FAILED - Password not given or incorrect

 LOGOFF SUCCESSFUL - Connection terminated successfully

To log on to the 9110T analyzer with SECURITY MODE feature enabled, type:

LOGON 940331

940331 is the default password. To change the default password, use the variable RS-232_PASS issued
as follows:

V RS-232_PASS=NNNNNN

Where N is any numeral between 0 and 9.