No response, Unstable zero and span – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 277

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

277

 If the response offset test functions for NO (NO OFFS) or NO

X

(NOX OFFS) are greater than

150 mV, a reaction cell contamination is indicated.

 Clean the reaction cell as described in Section 11.3.9.

12.5.2. NO RESPONSE

If the instrument shows no response (display value is near zero) even though sample gas is supplied
properly and the instrument seems to perform correctly.

1. Carry out an electrical test with the ELECTRICAL TEST procedure in the

diagnostics menu, see Section 12.7.12.2.

 If this test produces a concentration reading, the analyzer’s electronic signal

path is correct.

2. Carry out an optical test using the OPTIC TEST procedure in the diagnostics

menu, see Section 12.7.12.1.

 If this test results in a concentration signal, then the PMT sensor and the

electronic signal path are operating properly.

 If the 9110T passes both ETEST and OTEST, the instrument is capable of

detecting light and processing the signal to produce a reading.

 Therefore, the problem must be in the pneumatics or the ozone generator.

3. Check if the ozone generator is turned on.

 Usually, the analyzer issues a warning whenever the ozone generator is turned

off.

 Go to SETUP-MORE-DIAG-ENTR, then scroll to the OZONE GEN OVERRIDE

and see if it shows ON.

 If it shows OFF, turn it ON and EXIT the DIAG menu.

 If this is done and the ozone flow is correct, the analyzer should be properly

supplied with ozone unless the generator itself is broken.

4. Confirm the lack of response by supplying NO or NO2 span gas of about

80% of the range value to the analyzer.

5. Check the sample flow and ozone flow rates for proper values.

6. Check for disconnected cables to the sensor module.

7. If NO

2

signal is zero while NO signal is correct, check the NO/NOx valve

and the NO

2

converter for proper operation.

12.5.3. UNSTABLE ZERO AND SPAN

Leaks in the 9110T or in the external gas supply and vacuum systems are the most common source of
unstable and non-repeatable concentration readings.