Using the analog output test channel – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 269

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

269

12.2. USING THE ANALOG OUTPUT TEST CHANNEL

The signals available for output over the 9110T’s analog output channel can also be used as diagnostic
tools. See Section 5.9.2 for instruction on activating the analog output and selecting a function.

Table 12-3: Test Channel Outputs as Diagnostic Tools

TEST

CHANNEL

DESCRIPTION ZERO

FULL

SCALE

CAUSES OF EXTREMELY

HIGH / LOW READINGS

PMT

DETECTOR

The output of the PMT

detector converted to a 0

to 5 VDC scale.

0 mV

5000 mV

Failed PMT
PMT Temperature too High/Low
Bad PMT Preamp PCA
Failed HVPS
Misadjusted HVPS drive Voltage
Light Leak in reaction cell

OZONE

FLOW

The flow rate of O

3

through the analyzer as

measured by the O

3

flow

sensor

0

cm

3

/min

1000

cm

3

/min

Check for Gas Flow problems in the O

3

gas lines.

SAMPLE

FLOW

The calculated flow rate

for sample gas through

the analyzer.

0

cm

3

/min

1000

cm

3

/min

Check for Gas Flow problems in the sample gas lines.

SAMPLE

PRESSURE

The pressure of the

sample gas measured

upstream of the Auto

Zero Valve

0 In-Hg-A

40 In-Hg-A

Check for Gas Flow problems in the sample gas lines.

RCELL

PRESSURE

The pressure of gas

inside the reaction cell of

the sensor module

0 In-Hg-A

40 In-Hg-A

Check for Gas Flow problems in all gas lines.

RCELL TEMP

The temperature of gas

inside the reaction cell of

the sensor module

0

C 70

C

Same as RCELL TEMP WARNING in Table 12-1.

IZS TEMP

The temperature of the

permeation tube oven of

the optional internal span

gas generator.

0

C 70

C

Same as IZS TEMP WARNING in Table 12-1.

CONV TEMP

The temperature NO

2

NO converter

0 mV

5000 mV

Same as CONV TEMP WARNING in Table 12-1.

PMT TEMP

The temperature inside

PMT

0

C 50

C

Same as PMT TEMP WARNING in Table 12-1.

BOX TEMP

The temperature inside

the 9110T’s chassis

0

C 70

C

Same as BOX TEMP WARNING in Table 12-1.

HVPS

VOLTAGE

Represents the output

voltage of the PMT's high

voltage power supply

0 mV

5000 mV

Same as HVPSWARNING in Table 12-1.

Advertising