Model 9110T NOx Analyzer

Communications Setup and Operation

Teledyne Analytical Instruments

156

 1 - The HESSEN protocol will always use RANGE1 for this gas list

entry

 2 - The HESSEN protocol will always use RANGE2 for this gas list

entry

 3 - Not applicable to the 9110T analyzer.

GAS ID

An identification number assigned to a specific gas. The 9110T analyzer
is a multiple gas instrument that measures NO

x

, NO and NO

2

. Their ID

numbers are as follows:

NO

x

211

NO 212
NO

2

213

REPORT

States whether this list entry is to be reported or not reported when ever
this gas type or instrument is polled by the HESSEN network. If the list
entry is not to be reported this field will be blank. It’s default gas list
consists of only reads:

NOX, 0, 211, REPORTED

NO, 0, 212, REPORTED

NO2, 0, 213, REPORTED

These default settings cause the instrument to report the concentration value of the currently active
range. If you wish to have just concentration value stored for a specific range, this list entry should be
edited or additional entries should be added to the list.

EXAMPLE: Changing the above NO

x

gas list entry to read:

NOX, 2, 211, REPORTED

Would only record the last NO

x

reading that occurred while RANGE2 (HIGH) range was active.

EDITING OR ADDING HESSEN GAS LIST ENTRIES

To add or edit an entry to the Hessen Gas List, press: