Model 9110T NOx Analyzer

Setup Menu

Teledyne Analytical Instruments

132

5.9.4. TEST CHAN OUTPUT (SELECTING A TEST CHANNEL FUNCTION

FOR OUTPUT A4)

The test functions available to be reported are listed in Table 5-9:

Table 5-9: Test Channels Functions available on the 9110T’s Analog Output

TEST CHANNEL

DESCRIPTION

ZERO

FULL SCALE

NONE

TEST CHANNEL IS TURNED OFF

PMT DETECTOR

The output of the PMT detector
converted to a 0 to 5 VDC scale.

0 mV

5000 mV

1

OZONE FLOW

The flow rate of O

3

through the analyzer

as measured by the O

3

flow sensor.

0 cm

3

/min 1000

cm

3

/min

SAMPLE FLOW

The calculated flow rate for sample gas
through the analyzer.

0 cm

3

/min 1000

cm

3

/min

SAMPLE PRESSURE

The pressure of the sample gas
measured upstream of the Auto Zero
Valve.

0 Hg-In-A

40 "Hg-In-A

RCELL PRESSURE

The pressure of gas inside the reaction
cell of the sensor module.

0 Hg-In-A

40 Hg-In-A

RCELL TEMP

The temperature of gas inside the
reaction cell of the sensor module.

0

C 70

C

MANIFOLD TEMP

Not used in the Model 9110T.

IZS TEMP

The temperature of the permeation tube
oven of the optional internal span gas
generator.

0

C 70

C

CONV TEMP

The temperature NO

2

 NO converter.

0

C 500

C

PMT TEMP

The temperature inside PMT.

0

C 50

C

BOX TEMP

The temperature inside the 9110T’s
chassis.

0

C 70

C

HVPS VOLTAGE

Represents the output voltage of the
PMT's high voltage power supply.

0 mV

5000 mV

1

Once a function is selected, the instrument not only begins to output a signal on the analog output, but
also adds TEST to the list of test functions viewable via the front panel display.