Evaluating no, No converter performance, No converter efficiency, press – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 295

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

295

STEP 3:

To cause the analyzer to calculate and record the NO

2

 NO converter efficiency, press:

Check the calculated efficiency

gain factor.

If the gain factor is NOT

between 0.9600 and 1.0200,

the

NO

2

NO converter

needs to be replaced.

M-P CAL

CONVERTER EFFICIENCY MENU

NO2 CAL

SET

EXIT

Starting from

CONVERTER EFFICIENCY MENU

(see preceding steps)

M-P CAL

CE FACTOR:1000.0 Gain

1.

0

0

0

0

ENTR EXIT

M-P CAL

CONVERTER EFFICIENCY MENU

NO2

CAL

SET

EXIT

SAMPLE

NO2 STB=XX.X PPB

<TST TST>

SETUP

SAMPLE RANGE=500.0 PPB NOX= XXXX

< TST

TST >

ENTR SETUP

Set the Display to show

the NOX STB test

function.

This function calculates

the stability of the NO/NO

x

measurement.

Toggle TST> button until ...

Allow NO

2

gas of the proper concetration to enter

the sample port at the rear of the analyzer.

Wait until NOX STB

falls below 0.5 ppb.

This may take several

minutes.

SAMPLE

NO2 STB=XX.X PPB

<TST TST>

ENTR

SETUP

M-P CAL

CONVERTER EFICIENCY MENU

NO2 CAL

SET

EXIT

M-P CAL

CE FACTOR=0.9852 Gain

0.

8

8

5

2

ENTR EXIT

Toggle these

buttons to initialize

the converter

efficiency at 1.0000.

The analyzer

calculates the

converter’s

efficiency.

12.7.10.4. Evaluating NO

2

NO Converter Performance

If the converter appears to have performance problems (conversion efficiency is outside of allowed range
of 96-102%), check the following: