Set up for gpt multipoint calibration of the 9110t – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 229

Advertising
background image

Model 9110T NOx Analyzer

EPA Protocol Calibration

Teledyne Analytical Instruments

229

10.6.1. SET UP FOR GPT MULTIPOINT CALIBRATION OF THE 9110T

The following 9110T features must be activated or set before calibration.

 Calibrate

the

NO

2

 NO converter (see Section 9.1.4).

 Set the reporting ranges for Independent mode (see Section 5.4.3.2).

 Turn ON the automatic temperature/pressure compensation (TPC) feature (see Section 5.8).

 Set the units of measure to ppb (see Section 5.4.3.4).

Figure 10-1:

GPT Calibration System