Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

360

13.7.1.1. AC Configuration – Internal Pump (JP7)

If your 9110T includes an internal pump the following table, jumper set JP7 is used to configure the
power supplied to it as shown in Figure 13-27.

Table 13-6: AC Power Configuration for Internal Pumps (JP7)

LINE

POWER

LINE

FREQUENCY

JUMPER

COLOR

FUNCTION

JUMPER

BETWEEN

PINS

110VAC

115 VAC

60 HZ

WHITE

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

Connects pump pins 2 & 4 to Neutral

4 to 9

50 HZ

1

BLACK

Connects pump pin 3 to 110 / 115 VAC power line

2 to 7

Connects pump pin 3 to 110 / 115 VAC power line

3 to 8

Connects pump pins 2 & 4 to Neutral

4 to 9

220VAC

240 VAC

60 HZ

BROWN

Connects pump pins 3 and 4 together

1 to 6

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

50 HZ

1

BLUE

Connects pump pins 3 and 4 together

1 to 6

Connects pump pin 1 to 220 / 240VAC power line

3 to 8

1

A jumper between pins 5 and 10 may be present on the jumper plug assembly, but has no function on the Model 9110T.

110 VAC /115 VAC

220 VAC /240 VAC

1

2

3

4

5

6

7

8

9

10

1

2

3

4

5

6

7

8

9

10

May be present on 50 Hz version of

jumper set, but is not functional on

the T200

Figure 13-27: Pump AC Power Jumpers (JP7)