5 ppm, 500 cm – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 227

Advertising
background image

Model 9110T NOx Analyzer

EPA Protocol Calibration

Teledyne Analytical Instruments

227

Equation 10-7

 

75

.

2

NO

O

RC

NO

O

NO

STD

R

F

F

V

F

F

F

NO

P

Note If

T

R

is >2 minutes or if P

R

is <2.75 ppm min, changes in flow conditions

(FT, FO, FNO) or in the reaction cell volume (VRC), or both will
have to be made, and T

R

and P

R

will have to be recalculated.

9. After Equation 10-5 and Equation 10-6 are satisfied, calculate the diluent air flow

(F

D

):

Equation 10-8

NO

O

T

D

F

F

F

F

10.5.2.1. Example GPT Calculation

The following is an example calculation that can be used to determine whether an existing GPT
calibrator will meet the required conditions for a specific calibration. For this example, it is assumed that
only the volume of the reaction cell, V

RC

, and the concentration of the NO standard, [NO]

STD

, are known.

All flow settings (F

NO

, F

O

, F

T

, and F

D

) will be calculated. In many uses, these flow settings are known

and need only to be substituted in Equations 8-5 and 8-6 to verify the required conditions. Before doing
any calculations, the URL and flow demand of the analyzer being calibrated must be known. Operating
parameters are determined from the operations manual:

 Upper range limit = 0.5 ppm, and

 Flow demand = 500 cm3/min.

 Volume of calibrator reaction cell is determined by physical measurement: VRC = 180 cm3

 The concentration of the NO standard gas to be used is determined by reference against an NIST-

SRM SRM (Section 2.0.7, Q.A. Handbook):

[NO]

STD

= 50.5 ppm

1. Determine the minimum total flow (F

T

) required at the output manifold:

F

T

= 500 cm

3

/min (110/100) = 550 cm

3

/min

Because low flows are difficult to control and measure, it is often advantageous to set a higher total flow
than needed. In this example, we will set F

T

to 2750 cm

3

/min.

2. Determine the highest NO concentration, [NO]

OUT

, required at the output manifold,:

[NO]

OUT

= 0.5 ppm (90/100) = 0.45 ppm

3. Calculate the NO flow (F

NO

) required to generate the NO concentration [NO]

OUT

:

Advertising