Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 70

Advertising
background image

Model 9110T NOx Analyzer

Getting Started

Teledyne Analytical Instruments

70

Figure 3-26:

Rear Panel Layout with Internal Span Source (IZS) OPT 50G

INTERNAL SPAN GAS GENERATION

The primary component of the internal span option is a permeation tube containing liquid NO

2.

As zero

air is passed over a permeable membrane on the end of the tube, molecules of NO

2

slowly pass through

the membrane mixing with the zero air.

The resulting concentration of the NO

2

span gas is determined by three factors:

 Size of the membrane: The larger the area of the membrane, the more permeation occurs.

 Temperature of the NO

2

: Increasing the temperature of the permeation tube increases the pressure

inside the tube and therefore increases the rate of permeation.

 Flow rate of the zero air: If the previous two variables are constant, the permeation rate of the NO

2

into the zero air stream will be constant. Therefore, a lower flow rate of zero air produces higher
concentrations of NO

2

.

In the Model 9110T the permeation tube enclosure is heated to a constant 50° C (10° above the
maximum operating temperature of the instrument) in order to keep the permeation rate constant. A
thermistor measures the actual temperature and reports it to the CPU for control feedback.

The flow rate of zero air across the permeation tube is maintained at 50 ± 10 cm³/min by a critical flow
orifice located in the analyzer’s exhaust manifold.

NITRIC ACID AND THE CHEMISTRY OF NO2 PERMEATION TUBES

The reaction of H

2

O with NO

2

to form HNO

3

(nitric acid) takes place whenever water and NO

2

are

present in the same gas mixture. In the 9110T this is mitigated as much as possible by passing the air
supply for the span gas generator through a special dryer, however the permeable membrane of the NO

2

tube will still allow H

2

O from the ambient environment to slowly collect in the tube at increasingly

higher concentrations. Over time this results in the presence of HNO

3

in the permeation tube which is

exuded into the 9110T’s pneumatics along with NO

2

.

Advertising