Warning fault condition possible causes – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 265

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

265

WARNING FAULT

CONDITION

POSSIBLE

CAUSES

RCELL PRESS WARN

Sample Pressure is <15 in-Hg or >

35 in-Hg

Normally 29.92 in-Hg at sea level

decreasing at 1 in-Hg per 1000 ft of

altitude (with no flow – pump

disconnected).

If Sample Pressure is < 15 in-HG:

 Blocked Particulate Filter
 Blocked Sample Inlet/Gas Line
 Failed Pressure Senor/circuitry

If Sample Pressure is > 35 in-HG:

 Bad Pressure Sensor/circuitry
 Pressure too high at Sample Inlet.

RCELL TEMP WARNING

RCELL temperature is < 45

C or >

55

C.

Heater configured for wrong voltage type
Failed RCELL Temperature Sensor
Relay controlling the heater is not working
Failed Relay Board
I

2

C Bus

REAR BOARD NOT DET

Motherboard not detected on

power up.

This WARNING only appears on Serial I/O COMM Port(s) Front

Panel Display will be frozen, blank or will not respond.

Failure of Motherboard

RELAY BOARD WARN

The CPU cannot communicate with

the Relay Board.

I

2

C Bus failure

Failed Relay Board
Loose connectors/wiring

SAMPLE FLOW WARN

Sample flow rate is < 350 cc/min or >

600 cc/min.

Failed Sample Pump
Blocked Sample Inlet/Gas Line
Dirty Particulate Filter
Leak downstream of RCELL Critical Flow Orifice
Failed Sample Pressure Sensor
Failed Vacuum Pressure Sensor

SYSTEM RESET

The computer has rebooted.

This message occurs at power on.
If it is confirmed that power has not been interrupted:
Failed +5 VDC power
Fatal Error caused software to restart
Loose connector/wiring

Note

A failure of the analyzer’s CPU, motherboard or power supplies can result
in any or ALL of the above messages.