Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

294

Input the starting NO

2

concentration value into the 9110T by pressing:

Converter Efficiency Menu

SAMPLE

RANGE=500.0 PPB NO=XXXX

<TST TST>

CAL

SETUP

SAMPLE

RANGE TO CAL

LOW HIGH

ENTR

EXIT

M-P CAL

CONCENTRATON MENU

NOX

NO

CONV

EXIT

M-P CAL

CONVERTER EFICIENCY MENU

NO2

CAL

SET

EXIT

M-P CAL

NO2 CE CONC: 500.0 Conc

0

4

0

0.

0

0

ENTR EXIT

SAMPLE

RANGE=500.0 PPB NO=XXXX

<TST TST>

CAL

SETUP

This step only appears if the
analyzer’s reporting range is

set for AUTO range mode.

Select LOW and press ENTR.

Repeat entire procedure for

HIGH range.

The expected NO

2

span

concentration value defaults

to 400.0 Conc.

Make sure that you specify

the actual concentration value

of the NO

2

gas.

Toggle these buttons

to change this value to

the concentration of

the NO

2

gas being

used.

Use these buttons to

select the appropriate

range.

Repeat entire procedure

for each range.

SAMPLE

RANGE=500.0 PPB NO=XXXX

<TST TST> ZERO

CONC

SETUP