Input the starting no, Concentration value into the 9110t by pressing – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual
Page 294
Model 9110T NOx Analyzer
Troubleshooting & Service
Teledyne Analytical Instruments
294
Input the starting NO
2
concentration value into the 9110T by pressing:
Converter Efficiency Menu
SAMPLE
RANGE=500.0 PPB NO=XXXX
<TST TST>
CAL
SETUP
SAMPLE
RANGE TO CAL
LOW HIGH
ENTR
EXIT
M-P CAL
CONCENTRATON MENU
NOX
NO
CONV
EXIT
M-P CAL
CONVERTER EFICIENCY MENU
NO2
CAL
SET
EXIT
M-P CAL
NO2 CE CONC: 500.0 Conc
0
4
0
0.
0
0
ENTR EXIT
SAMPLE
RANGE=500.0 PPB NO=XXXX
<TST TST>
CAL
SETUP
This step only appears if the
analyzer’s reporting range is
set for AUTO range mode.
Select LOW and press ENTR.
Repeat entire procedure for
HIGH range.
The expected NO
2
span
concentration value defaults
to 400.0 Conc.
Make sure that you specify
the actual concentration value
of the NO
2
gas.
Toggle these buttons
to change this value to
the concentration of
the NO
2
gas being
used.
Use these buttons to
select the appropriate
range.
Repeat entire procedure
for each range.
SAMPLE
RANGE=500.0 PPB NO=XXXX
<TST TST> ZERO
CONC
SETUP