Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 175

Advertising
background image

Model 9110TH NOx Analyzer

Data Acquisition System (DAS) and APICOM

Teledyne Analytical Instruments

175

The SAMPLE PERIOD and REPORT PERIOD intervals are synchronized to the beginning and end
of the appropriate interval of the instruments internal clock.

 If

SAMPLE PERIOD is set for one minute the first reading would occur at the beginning of the next

full minute according to the instrument’s internal clock.

 If

the

REPORT PERIOD is set for of one hour, the first report activity would occur at the beginning of

the next full hour according to the instrument’s internal clock.

EXAMPLE:

Given the above settings, if DAS parameters are activated at 7:57:35 the first sample would occur at 7:58
and the first report would be calculated at 8:00 consisting of data points for 7:58. 7:59 and 8:00.

During the next hour (from 8:01 to 9:00), the instrument will take a sample reading every minute and
include 60 sample readings.

7.1.3.5. Report Periods in Progress when Instrument Is Powered Off

If the instrument is powered off in the middle of a REPORT PERIOD, the samples accumulated during
that period are lost. Once the instrument is turned back on, the DAS restarts taking samples and