Teledyne 9110T - Nitrogen Oxides Analyzer User Manual
Page 121
Model 9110TH NOx Analyzer
Setup Menu
Teledyne Analytical Instruments
121
5.9.3.4. Automatic Group Calibration of the Analog Outputs
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Manual calibration should be used for any analog output set for a 0.1V
output range or in cases where the outputs must be closely matched to
the characteristics of the recording device. (See Sections 5.9.3.2, 5.9.3.3,
and 5.9.3.6).
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
Before performing this procedure, ensure that the AUTO CAL for each
analog output is enabled. (See Section 5.9.3.3).
To calibrate the outputs as a group with the AOUTS CALIBRATION command, select the ANALOG
I/O CONFIGURATION submenu (see Figure 5-5) then press:
5.9.3.5. Automatic Individual Calibration of the Analog Outputs
To use the AUTO CAL feature to initiate an automatic calibration for an individual analog output, select
the ANALOG I/O CONFIGURATION submenu (see Figure 5-5) then press: