Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 287

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

287

1. Use the Signal I/O function (see Section 12.1.3 and Appendix A) to view the value

of REF_4096_MV and REF_GND.

 If both are within 3 mV of nominal (4096 and 0), and are stable, ±0.2 mV then

the basic A/D is functioning properly. If not then the motherboard is bad.

2. Choose a parameter in the Signal I/O function list (see Section 12.1.3) such as

OZONE_FLOW .

 Compare this voltages at its origin (see the interconnect drawing and

interconnect list in Appendix D) with the voltage displayed through the signal I/O
function.

 If the wiring is intact but there is a large difference between the measured and

displayed voltage (±10 mV) then the motherboard is bad.

12.7.6.3. Status Outputs

V

+DC Gnd

Figure 12-5:

Typical Set Up of Status Output Test

To test the status output electronics:

1. Connect a jumper between the “D" pin and the “” pin on the status output

connector.

2. Connect a 1000 ohm resistor between the “+” pin and the pin for the status output

that is being tested.

3. Connect a voltmeter between the “” pin and the pin of the output being tested.

4. Under the DIAG Signal I/O menu (see Section 12.1.3), scroll through the inputs

and outputs until you get to the output in question.