Model 9110T NOx Analyzer

Calibration Procedures

Teledyne Analytical Instruments

218

EXIT returns

to the SETUP

Menu.

SETUP

DURATION:15.0 MINUTES


<SET SET>

EDIT

EXIT

SETUP

DURATION 15.0MINUTES


3 0 .0

ENTR

EXIT

SETUP

DURATION:30.0 MINUTES


<SET

SET>

EDIT

EXIT

Toggle buttons to set

duration for each iteration

of the sequence:

Set in Decimal minutes

from 0.1 – 60.0.

SETUP

CALIBRATE: OFF


<SET SET>

EDIT

EXIT

SETUP

CALIBRATE: OFF


ON

ENTR

EXIT

SETUP X.X

CALIBRATE: ON


<SET SET> EDIT

EXIT

Toggle button

Between Off and

ON.

SETUP X.X

SEQ 2) ZERO–SPAN, 2:00:30


PREV NEXT

MODE

SET

EXIT

Display show:

SEQ 2) ZERO–SPAN, 2:00:30


Sequence

Delta Time

MODE Delta Days

CONTINUED FROM PREVIOUS PAGE

DELTA TIME

Note

If at any time an unallowable entry is selected (Example: Delta Days > 367)
the ENTR button will disappear from the display.