Warning fault condition possible causes – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 264

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

264

Table 12-1: Front Panel Warning Messages

WARNING FAULT

CONDITION

POSSIBLE

CAUSES

AZERO WARN

XXX.X MV

Auto-zero reading above 200 mV.

Value shown in message indicates

auto-zero reading at time warning

was displayed.

AZERO valve not working
Valve control driver failed
Bad Relay PCA
Failed +12 VDC power supply
Gas leak across AZERO Valve ports
Dirty Reaction Cell
O

3

flow problem to RCELL

BOX TEMP WARNING

Box Temp is < 7

C

or > 48

C.

Box Temperature typically runs ~7

C warmer than ambient

temperature

Poor/blocked ventilation to the analyzer
Stopped Exhaust-Fan
Ambient Temperature outside of specified range

CANNOT DYN SPAN

Dynamic Span operation failed.

Measured concentration value is too high or low
Concentration Slope value to high or too low

CANNOT DYN ZERO

Dynamic Zero operation failed.

Measured concentration value is too high
Concentration Offset value to high

CONFIG INITIALIZED

Configuration and Calibration data

reset to original Factory state.

Failed Disk on Module
User erased data

CONV TEMP WARNING

NO

2

NO Converter temperature <

305

C or > 325C.

Heater configured for wrong voltage type
Failed converter Temperature Sensor
Relay controlling the Heater is not working
Failed Relay Board

DATA INITIALIZED

Data Storage in DAS was erased.

Failed Disk-on-Module
User cleared data.

HVPS WARNING

High voltage power supply output

outside of warning limits specified by

HVPS_SET

variable.

No +15 VDC power supply to Preamplifier PCA
Drive voltage not adjusted properly
Failed PMT Preamplifier PCA
Dirty reaction cell
Bad pneumatic flow

IZS TEMP WARNING

Permeation tube oven temperature is

< 45

C or > 55C.

Heater configured for wrong voltage type
Failed permeation tube Temperature Sensor
Relay controlling the Heater is not working
Failed Relay Board

OZONE FLOW

WARNING

O

3

flow rate is < 50 cc/min or >

150 cc/min.

Failed Sample Pump
Blocked O

3

dryer

Blocked inlet/outlet to O

3

purifier

Dirty O

3

dryer DFU

Leak downstream of RCELL
Failed O

3

Flow Sensor

OZONE GEN OFF

Ozone generator is off. This is the

only warning message that

automatically clears itself. It clears

itself when the ozone generator is

turned on.

O

3

generator override is turned ON.

Electrical connection between motherboard and generator is

faulty.

Bad +15VDC power supply

PMT TEMP WARNING

Sample temperature is < 5

C or >

12

C.

PMT fan not operating
Failed PMT Temperature Sensor
TEC not functioning
Failed PMT Preamp PCA