Short calibration checks, Zero/span check procedures – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 233

Advertising
background image

Model 9110T NOx Analyzer

EPA Protocol Calibration

Teledyne Analytical Instruments

233

Table 10-3: Activity Matrix for Data Quality

Characteristic

Acceptance Limits

Frequency and Method

of Measurement

Action if Requirements

are not Met

Shelter
temperature

Mean temperature between 22

C

and 28

C (72 and 82 F), daily

fluctuations not greater than

 2C

Check thermograph chart
weekly for variations greater
than ± 2

C (4F)

Mark strip chart for the
affected time period.
Repair or adjust
temperature control.

Sample
introduction
system

No moisture, foreign material,
leaks, obstructions; sample line
connected to manifold

Weekly visual inspection

Clean, repair, or replace as
needed.

Recorder

Adequate ink & paper
Legible ink traces
Correct chart speed and range
Correct time

Weekly visual inspection

Replenish ink and paper
supply.
Adjust time to agree with
clock; note on chart.

Analyzer
operational
settings

TEST measurements at nominal
values
9110T in SAMPLE mode

Weekly visual inspection

Adjust or repair as needed.

Analyzer
operational
check

Zero and span within tolerance
limits as described in Section 12 of
the Q.A. Handbook

6

Level 1 zero/span every
2 weeks; Level 2 between
Level 1 checks at frequency
desired by user

Find source of error and
repair.
After corrective action, re-
calibrate analyzer.

Precision check Assess precision as described in

Sections 15&18 of the Q.A.
Handbook

6

Every 2 weeks, Subsection
3.4.3 (Ibid.)

Calc, report precision,
Section 12 of the Q.A.
Handbook

6

.

10.8.2. SHORT CALIBRATION CHECKS

A system of Level 1 and Level 2 zero/span checks (Table 10-2) is recommended. These checks must be
conducted in accordance with the specific guidance given in Section 12 of the Q.A. Handbook

6

. Level 1

zero and span checks must be conducted every two weeks. Level 2 checks should be conducted in
between the Level 1 checks at a frequency desired by the user. Span concentrations for both levels
should be between 70 and 90% of the measurement range.

Zero and span data are to be used to:

 Provide data to allow analyzer adjustment for zero and span drift;

 Provide a decision point on when to calibrate the analyzer;

 Provide a decision point on invalidation of monitoring data.

These items are described in detail in Sections 15 & 18 of the Q.A. Handbook

6

. Refer to Section 11 of this

manual if the instrument is not within the allowed margins. We recommend using APICOM and the DAS
for analysis and documentation of zero/span check data.

10.8.3. ZERO/SPAN CHECK PROCEDURES

The Zero and span calibration can be checked in a variety of different ways. They include:

 Manual zero/span checks can be done from the front panel touchscreen. The procedure is in

Section 9.2.2 of this manual.

 Automatic zero/span checks can be performed every night. See Section 9.5 of this manual for setup

and operation procedures.