Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 44

Advertising
background image

Model 9110T NOx Analyzer

Getting Started

Teledyne Analytical Instruments

44

Figure 3-9:

Current Loop Option Installed on the Motherboard

CONVERTING CURRENT LOOP ANALOG OUTPUTS TO STANDARD VOLTAGE OUTPUTS

To convert an output configured for current loop operation to the standard 0 to 5 VDC output operation:

1. Turn off power to the analyzer.

8. If a recording device was connected to the output being modified, disconnect it.

9. Remove the top cover.

 Remove the set screw located in the top, center of the rear panel.

 Remove the screws fastening the top cover to the unit (one per side).

 Slide the cover back and lift the cover straight up.

10. Remove the screw holding the current loop option to the motherboard.

11. Disconnect the current loop option PCA from the appropriate connector on the

motherboard (see Figure 3-9).

12. Each connector, J19 and J23, requires two shunts. Place one shunt on the two left

most pins and the second shunt on the two adjacent pins (see Figure 3-9).

13. Reattach the top case to the analyzer.

Advertising