Epa equivalency designation – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 28

Advertising
background image

Model 9110T NOx Analyzer

Specifications, Approvals, & Compliance

Teledyne Analytical Instruments

28

Parameter

Specification

External Pump Pack: 15 lbs (7 kg)

Operating Temperature Range

5 - 40

C (with EPA equivalency)

Humidity Range

0-95% RH non-condensing

Environmental Conditions

Installation Category (Over voltage Category) II Pollution Degree 2

1

As defined by the US EPA.

2

Defined as twice the zero noise level by the US EPA.

2.2. EPA EQUIVALENCY DESIGNATION

TAI’s 9110T nitrogen oxides analyzer is designated as a reference method (Number RFNA-1104-099)
for NO

2

measurement, as defined in 40 CFR Part 53, when operated under the following conditions:

 Range: Any full-scale range between 0-0.05 and 0-1.0 ppm (parts per million).

 Ambient temperature range of 5 to 40°C.

 With 1-micron PTFE filter element installed in the internal filter assembly.

 Equipped with ozone supply air filter

 Gas flow supplied by External vacuum pump capable of 10 in-Hg-A at 2 standard liters per minute

(slpm) or better.

 Software Settings, see Table 2-2:

Table 2-2: Software Settings for EPA Equivalence

Parameter Setting

Dynamic Zero

OFF or ON

Dynamic Span

OFF

CAL-on-NO

2

OFF

Dilution Factor

1.0 or OFF

Temp/Pres compensation

ON

AutoCal

ON or OFF

Independent range

ON or OFF

Auto range

ON or OFF

Under the designation, the Analyzer may be operated with or without the following options:

 Rack mount with or without slides.

 Rack mount for external pump.

 4-20mA isolated analog outputs.

 Zero/Span Valves option.

 Internal Zero/Span (IZS) option with:

 NO

2

permeation tube - 0.4 ppm at 0.7 liter per minute; certified/uncertified.

 NO

2

permeation tube - 0.8 ppm at 0.7 liter per minute; certified/uncertified.