Model 9110T NOx Analyzer
Troubleshooting & Service
Teledyne Analytical Instruments
299
12.7.11.1. Simplified GPT Data Sheet
Line # TEST
RESULT
1
LEAK-CHECK (WHEN HOT)
YES / NO
2
NO
x
RESPONSE (MOLY BYPASSED)
__________
3
NO
x
RESPONSE (MOLY IN-LINE)
__________
4
OUT-GASSING (NO – NOX)
__________ (>-5 ppb, <5 ppb)
5 (
NO
x
ORIG
)
(
NO
x
mode, O
3
off)
__________ ppb
6 (
NO
x
REM
)
(
NO
x
mode, O
3
on)
__________ ppb
7
NO
x
LOSS
__________
(
A
)
(<4%
of
NO
x
ORIG
:
for 450PP 4% is 18 ppb)
8 (NO
ORIG
)
(NO mode, O3 off)
__________ ppb
9 (NO
REM
)
(NO mode, O3 on)
__________ ppb
10
NO
2
__________ (
B
) (>300ppb)
11 Efficiency LOSS [ (
A
/
B
) x 100 ] = [ ( ____
A
/ ____
B
) x 100 ] = ____
C
%
12 Total Conv Eff [ 100% –
C
] = 100% - ____
C
= _____
D
% ( > 96%)