Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 299

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

299

12.7.11.1. Simplified GPT Data Sheet

Line # TEST

RESULT

1

LEAK-CHECK (WHEN HOT)

YES / NO

2

NO

x

RESPONSE (MOLY BYPASSED)

__________

3

NO

x

RESPONSE (MOLY IN-LINE)

__________

4

OUT-GASSING (NO – NOX)

__________ (>-5 ppb, <5 ppb)

5 (

NO

x

ORIG

)

(

NO

x

mode, O

3

off)

__________ ppb

6 (

NO

x

REM

)

(

NO

x

mode, O

3

on)

__________ ppb

7

NO

x

LOSS

__________

(

A

)

(<4%

of

NO

x

ORIG

:

for 450PP 4% is 18 ppb)

8 (NO

ORIG

)

(NO mode, O3 off)

__________ ppb

9 (NO

REM

)

(NO mode, O3 on)

__________ ppb

10

NO

2

__________ (

B

) (>300ppb)

11 Efficiency LOSS [ (

A

/

B

) x 100 ] = [ ( ____

A

/ ____

B

) x 100 ] = ____

C

%

12 Total Conv Eff [ 100% –

C

] = 100% - ____

C

= _____

D

% ( > 96%)