Calibration problems, Negative concentrations – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 276

Advertising
background image

Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

276

12.5. CALIBRATION PROBLEMS

This section describes possible causes of calibration problems.

12.5.1. NEGATIVE CONCENTRATIONS

Negative concentration values can be caused for several reasons:

 A slight, negative signal is normal when the analyzer is operating under zero gas and the signal is

drifting around the zero calibration point.

 This is caused by the analyzer’s zero noise and may cause reported concentrations to be

negative for a few seconds at a time down to -20 ppb, but should randomly alternate with
similarly high, positive values.

 The 9110T has a built-in Auto Zero function, which should take care of most of these

deviations from zero, but may yield a small, residual, negative value.

 If larger, negative values persist continuously, check if the Auto Zero function was

accidentally turned off using the remote variables in Appendix A-2.

 In this case, the sensitivity of the analyzer may be drifting negative.

 A corruption of the Auto Zero filter may also cause negative concentrations.

 If a short, high noise value was detected during the Auto Zero cycle, that higher reading will

alter the Auto Zero filter value.

 As the value of the Auto Zero filter is subtracted from the current PMT response, it will

produce a negative concentration reading.

 High Auto Zero readings can be caused by

 a leaking or stuck Auto Zero valve (replace the valve),

 by an electronic fault in the preamplifier causing it to have a voltage on the

PMT output pin during the Auto Zero cycle (replace the preamplifier),

 by a reaction cell contamination causing high background (>40 mV) PMT

readings (clean the reaction cell),

 by a broken PMT temperature control circuit, allowing high zero offset

(repair the faulty PMT cooler). After fixing the cause of a high Auto Zero
filter reading, the 9110T will take 15 minutes for the filter to clear itself, or

 by an exhausted chemical in the ozone cleanser (see Section 11.3.3).

 Calibration error is the most likely explanation for negative concentration values.

 If the zero air contained some NO or NO

2

gas (contaminated zero air or a worn-out zero air

scrubber) and the analyzer was calibrated to that concentration as “zero”, the analyzer may
report negative values when measuring air that contains little or no NO

x

.

 The same problem occurs, if the analyzer was zero-calibrated using zero gas that is

contaminated with ambient air or span gas (cross-port leaks or leaks in supply tubing or
user not waiting long enough to flush pneumatic systems).

Advertising