Model 9110TH NOx Analyzer

Getting Started

Teledyne Analytical Instruments

79

 Cal gas will be supplied through the SAMPLE gas inlet on the back of the analyzer and;

 The pneumatic setup matches that described in Section 3.3.2.

VERIFYING THE REPORTING RANGE SETTINGS

While it is possible to perform the following procedure with any range setting we recommend that you
perform this initial checkout using following reporting range settings:

 Unit of Measure: PPB

 Reporting Range: 500 ppb

 Mode Setting: SNGL

While these are the default setting for the 9110T analyzer, it is recommended that you verify them before
proceeding with the calibration procedure, by pressing the following menu button sequence: