Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

361

13.7.1.2. AC Configuration – Standard Heaters (JP2)

Power configuration for the AC the standard heaters is set using Jumper set JP2 (see Figure 13-28 for the
location of JP2).

Table 13-7: Power Configuration for Standard AC Heaters (JP2)

LINE VOLTAGE

JUMPER

COLOR

HEATER(S)

JUMPER

BETWEEN

PINS

FUNCTION

110 VAC / 115 VAC

50Hz & 60 Hz

WHITE

Reaction Cell / Sample

Chamber Heaters

1 to 8

Common

2 to 7

Neutral to Load

4 to 9

Neutral to Load

Moly Converter

3 to 10

Common

4 to 9

Neutral to Load

6 to 11

Neutral to Load

220 VAC / 240 VAC

50Hz & 60 Hz

BLUE

Reaction Cell / Sample

Chamber Heaters

1 to 7

Load

Moly Converter

3 to 9

Load

1

2

3

4

5

6

7

8

9

10

11

12

Reaction Cell

Heaters

NO

2

NO

Converter Heaters

110 VAC /115 VAC

1

2

3

4

5

6

7

8

9

10

11

12

220 VAC / 240 VAC

Reaction Cell

Heaters

NO

2

NO

Converter Heaters

Figure 13-28: Typical Set Up of AC Heater Jumper Set (JP2)

13.7.1.3. AC Configuration – Heaters for Option Packages (JP6)

The IZS valve option includes an AC heaters that maintain an optimum operating temperature for key
components of those options. Jumper set JP6 is used to connect the heaters associated with those options
to AC power. Since these heaters work with either 110/155 VAC or 220/240 VAC, there is only one
jumper configuration.