Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

348

D7 (Green) – Optional Zero/Span Valve

D11 (Green) – Optional Dual Span Select Valve

D12 (Green) – Optional Pressurized Span Shutoff Valve

D3 (Yellow) NO

2

NO Converter Heater

D5 (Yellow) – Optional Internal Span Gas Gen Heater

D8 (Green) – Optional Sample/Cal Valve

D1 (RED)

Watchdog Indicator

D2 (Yellow) Reaction Cell Heater

D9 (Green) – AutoZero Valve

D10 (Green) – NO/NO

x

Valve

D13 (Green) – Optional Pressurized Zero Shutoff Valve

Figure 13-17: Status LED Locations – Relay PCA

13.3.4.2. Watchdog Circuitry

The most important of the status LED’s on the relay board is the red I

2

C bus watch-dog LED. It is

controlled directly by the analyzer’s CPU over the I

2

C bus. Special circuitry on the relay PCA watches

the status of D1. Should this LED ever stay ON or OFF for 30 seconds, indicating that the CPU or I

2

C

bus has stopped functioning, this Watchdog Circuit automatically shuts all valves and turns off all
heaters.

13.3.4.3. Valve Control

The relay board also hosts two valve driver chips, each of which can drive up four valves. The main
valve assembly in the 9110T is the NO/NO

X

- Auto-zero solenoid valve component mounted right in

front of the NO

2

converter housing (see Figure 3-5).

 These two valves are actuated with 12 V supplied from the relay board and under the control of the

CPU through the I

2

C bus.

Additional valve sets also controlled by the CPU via the I

2

C bus and the relay PCA can be included in

the 9110T (see Table 1-1 and Sections 3.3.2.3, 3.3.2.4, and 3.3.2.5 for descriptions of these valve sets).