Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

345

13.3.3.8. Power-Up Circuit

This circuit monitors the +5V power supply during start-up and sets the analog outputs, external digital
I/O ports, and I

2

C circuitry to specific values until the CPU boots and the instrument software can

establish control.

13.3.4. RELAY PCA

The CPU issues commands via a series of relays and switches located on a separate printed circuit
assembly, called the relay PCA, to control the function of key electromechanical devices such as heaters
and valves. The relay PCA receives instructions in the form of digital signals over the I

2

C bus, interprets

these digital instructions and activates its various switches and relays appropriately.

The relay PCA is located in the right-rear quadrant of the analyzer and is mounted vertically on the
backside of the same bracket as the instrument’s DC power supplies.

Power

Connection

for DC

Heaters

Status LED’s

(D2 through D16)

DC Power Supply

Test Points

Watchdog

Status LED (D1)

(JP5)

Thermocouple

Configuration

Jumpers

Thermocouple

Signal Output

I

2

C

Connector

Valve Control

Drivers

Pump Power

Output

Pump AC

Configuration

Jumper

AC Power

IN

DC Power

Distribution

Connectors

Valve Control

Connector

TC1 Input

NO

2

NO Converter

Temp Sensor

Connector for

AC Relays

K1 & K2

Connector for AC Relays K4 & K5

Heater AC Power

Configuration

Jumpers

JP2

JP6

J15

J16

R16

J17

J2

JP7

J18

TP1

TP2

TP3

TP4

TP5

TP6

TP7

J21
J19
J14

U5 U6

J4

J12

J11

J10

J9

J8

J7

J5

J1

3

J3

AC Relay K1

(Reaction Cell Heater)

AC Relay K2

(NO

2

NO Converter Heater)

AC Relay K4

(OPT Internal Span Gen Heater)

Figure 13-15: Relay PCA Layout (P/N 045230100)

CAUTION

E

LECTRICAL

S

HOCK

H

AZARD

Only those relays actually required by the configuration of the 9110T are populated.

A protective retainer plate is installed over the ac power relay to keep them securely

Advertising