Model 9110T NOx Analyzer

Getting Started

Teledyne Analytical Instruments

60

Figure 3-18:

Gas Line Connections from Bottled Span Gas – Basic 9110T Configuration

For the 9110T nitrogen oxides analyzer in its basic configuration, attach the following pneumatic lines:

SAMPLE GAS SOURCE

Connect a sample gas line to the SAMPLE inlet

 Use PTFE tubing; minimum OD ¼”

 Sample Gas pressure must equal ambient atmospheric pressure (1.0 psig)

 In applications where the sample gas is received from a pressurized manifold and the analyzer is

not equipped with one of the 9110T’s pressurized span options, a vent must be placed on the
sample gas line. This vent line must be:

 No more than 10 meters long.

 Vented outside the shelter or immediate area surrounding the instrument.

CALIBRATION GAS SOURCES

CAL GAS & ZERO AIR SOURCES: The source of calibration gas is also attached to the SAMPLE

inlet, but only when a calibration operation is actually being performed.

 Use PTFE tubing; minimum OD ¼”.