Model 9110T NOx Analyzer

Troubleshooting & Service

Teledyne Analytical Instruments

284

Table 12-6:

DC Power Test Point and Wiring Color Codes

NAME TEST

POINT#

COLOR DEFINITION

DGND

1

Black Digital

ground

+5V

2

Red

AGND

3

Green Analog

ground

+15V

4

Blue

-15V

5

Yellow

+12R

6

Purple

12 V return

(ground) line

+12V

7

Orange

TP1 TP2 TP3 TP4 TP5 TP6 TP7
DGND +5V AGND +15V -15V +12R 12V

Figure 12-4:

Location of DC Power Test Points on Relay PCA

A voltmeter should be used to verify that the DC voltages are correct per the values in the table below,
and an oscilloscope, in AC mode, with band limiting turned on, can be used to evaluate if the supplies
are producing excessive noise (> 100 mV p-p).

Table 12-7: DC Power Supply Acceptable Levels

POWER

SUPPLY

VOLTAGE CHECK

RELAY BOARD TEST POINTS

MIN V

MAX V

FROM

Test Point

TO

Test Point

NAME

#

NAME

#

PS1 +5

DGND

1

+5 2

+4.85

+5.25

PS1 +15

AGND

3

+15 4

+13.5

+16.0

PS1 -15

AGND

3

-15V

5

-13.5

-16.0

PS1 AGND

AGND

3

DGND 1

-0.05

+0.05

PS1 Chassis

DGND

1

Chassis

N/A

-0.05

+0.05

PS2 +12

+12V

Ret

6

+12V

7

+11.8

+12.5

PS2 DGND

+12V

Ret

6

DGND 1

-0.05

+0.05