Other quality assurance procedures, F ] no [ * f + ] [no - ] [no = ] no, Summary of quality assurance checks – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 232

Advertising
background image

Model 9110T NOx Analyzer

EPA Protocol Calibration

Teledyne Analytical Instruments

232

4. While maintaining all other conditions, adjust the ozone generator to obtain several

other concentrations of NO

2

evenly spaced between the 80% URL point and the

zero point. Record the NO, NO

X

, and NO

2

concentrations for each additional point.

5. Calculate the resulting NO

2

concentrations as follows:

Equation

10-10

F

]

NO

[

*

F

+

]

[NO

-

]

[NO

=

]

NO

[

T

IMP

2

NO

REM

ORIG

OUT

2

Where [NO]

ORIG

is the NO concentration before the GPT ozone is turned on, and [NO]

REM

is the NO

remaining after GPT.

Plot the NO

2

concentration output by the instrument on the y-axis against the generated NO

2

[NO

2

]

OUT

on the x-axis. The plot should be a straight line within the ± 2% linearity criteria given for the NO

x

and

NO channels. If the plot is not linear, the most likely cause is that the converter needs replacing. See
Section 9.1.4 on NO

2

converter efficiency and Section 11.3.8 for changing the converter if needed.

10.8. OTHER QUALITY ASSURANCE PROCEDURES

Precision is determined by a one-point check at least once every two weeks. Accuracy is determined by a
three-point audit once each quarter.

Essential to quality assurance are scheduled checks for verifying the operational status of the monitoring
system. The operator should visit the site at least once each week. Every two weeks a Level 1 zero and
span check must be made on the analyzer. Level 2 zero and span checks should be conducted at a
frequency desired by the user. Definitions of these terms are given in Table 10-2.

10.8.1. SUMMARY OF QUALITY ASSURANCE CHECKS

The following items should be checked on a regularly scheduled basis to assure high quality data from
the 9110T. See Table 10-3 for a summary of activities; also the QA Handbook should be checked for
specific procedures.

Advertising