Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

334

Figure 13-11: Semi-Permeable Membrane Drying Process

The process by which the water vapor molecules are collected and transported through Nafion

®

material

is called per-evaporation and is driven by the humidity gradient between the inner and outer tubes as
well as the flow rates and pressure difference between inner and outer tubing. Unlike micro-porous
membrane permeation, which transfers water through a relatively slow diffusion process, per-
evaporation is a simple kinetic reaction. Therefore, the drying process occurs quickly, typically within
milliseconds.

Because this chemical reaction is based on hydrogen bonds between the water molecule and the Nafion

®

material most other chemical components of the gas to be dried are usually unaffected. Specifically, the
gases of interest for the 9110T, NO and NO

2

, do not get absorbed and pass the dryer unaltered.

On the other hand, other small polar gases that are capable of hydrogen bonds such as ammonia (NH

3

)

can be absorbed this way, too. This is an advantage since gases such as NH

3

can cause interference for

the measurement of NO

x

, NO and NO

2

(see Table 13-1).

Figure 13-12: 9110T Perma Pure

®

Dryer