Model 9110TH NOx Analyzer

Calibration Procedures

Teledyne Analytical Instruments

203

9.3.2. SETUP FOR CALIBRATION WITH THE INTERNAL SPAN GAS

GENERATOR

Connect the sources of zero air and span gas as shown in Figure 9-4.

Figure 9-4:

Pneumatic Connections for Manual Calibration/Checks with the Internal Span Gas
Generator

9.3.3. CAL ON NO

2

FEATURE

When using the IZS option to calibrate the 9110T, the analyzer’s CAL_ON_NO

2

feature must be turned

on. This feature enables a continuous zero gas flow across the IZS permeation tube and through the NO

2

converter. It also programs the analyzer to use the NO output from the NO

2

converter to calibrate the

span value of both NO and NO

X

.

Note

This feature should only be enabled when a span calibration or calibration
check is performed.

While CAL_ON_NO

2

is enabled, the NO

2

concentration will always be reported as zero. This is

because the gas is continuously routed through the NO

2

converter and the analyzer’s firmware simulates

calibration with NO gas.

Table 9-1: IZS Option Valve States with CAL_ON_NO

2

Turned ON

Valve

Condition

Valve Port Connections

Sample/Cal

Open to zero/span valve

1  2

Zero/Span

Open to SPAN GAS inlet

1  2

NO/NO

x

Valve

Open to NO

2

converter

1  2

Auto Zero Valve

Cycles normally

N/A

Since the instrument sees the same concentration of NO during both NO and NO

X

cycles, it reports an

NO

2

concentration of zero.

TO turn the CAL_ON_NO

2

feature ON/OFF, press: