Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 123

Advertising
background image

Model 9110TH NOx Analyzer

Setup Menu

Teledyne Analytical Instruments

123

5.9.3.6. Manual Calibration of the Analog Outputs Configured for Voltage Ranges

For highest accuracy, the voltages of the analog outputs can be manually calibrated.

Note

The menu for manually adjusting the analog output signal level will only
appear if the AUTO-CAL feature is turned off for the channel being
adjusted. (See Section 5.9.3.3).

Calibration is performed with a voltmeter connected across the output terminals and by changing the
actual output signal level using the front panel buttons in 100, 10 or 1 count increments. See Figure 3-8
for pin assignments and diagram of the analog output connector.

V

+DC Gnd

Figure 5-6:

Setup for Checking / Calibrating DCV Analog Output Signal Levels

Table 5-7: Voltage Tolerances for the TEST CHANNEL Calibration

FULL

SCALE

ZERO

TOLERANCE

SPAN VOLTAGE

SPAN

TOLERANCE

MINIMUM

ADJUSTMENT

(1 count)

0.1 VDC

±0.0005V

90 mV

±0.001V

0.02 mV

1 VDC

±0.001V

900 mV

±0.001V

0.24 mV

5 VDC

±0.002V

4500 mV

±0.003V

1.22 mV

10 VDC

±0.004V

4500 mV

±0.006V

2.44 mV