Checking for pneumatic leaks – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 258

Advertising
background image

Model 9110T NOx Analyzer

Instrument Maintenance

Teledyne Analytical Instruments

258

C

AUTION

Q

UALIFIED

P

ERSONNEL

O

NLY

This procedure is carried out with the analyzer running and its cover removed.

1. Scroll the front panel display to show then test function to PMT.

2. Supply zero gas to the analyzer.

3. With the instrument still running, carefully remove the analyzer cover.

WARNING

E

LECTRICAL

S

HOCK

H

AZARD

Do NOT touch any of the inside wiring with the metal cover or with your body.
Do NOT drop screws or tools into a running analyzer.

4. Shine a powerful flashlight or portable incandescent light at the inlet and outlet

fitting and at all of the joints of the reaction cell as well as around the PMT housing.

 The PMT value should not respond to the light, the PMT signal should remain

steady within its usual noise floor.

5. If there is a PMT response to the external light, symmetrically tighten the reaction

cell mounting screws or replace the 1/4” vacuum tubing with new, black PTFE
tubing (this tubing will fade with time and become transparent).

Note

Often, light leaks are also caused by O-rings being left out of the
assembly.

6. If, during this procedure, the black PMT housing end plate for the Sensor Assembly

is removed, ensure to replace the 5 desiccant bags inside the housing.

7. Carefully replace the analyzer cover. If tubing was changed, carry out a pneumatic

leak check.

11.3.12. CHECKING FOR PNEUMATIC LEAKS

C

AUTION

-

T

ECHNICAL

I

NFORMATION

Do not exceed 15 psi when pressurizing the system during either Simple or

Detailed checks.

11.3.12.1. Simple Vacuum Leak and Pump Check

Leaks are the most common cause of analyzer malfunction. This section presents a simple leak check,
whereas the next section details a more thorough procedure. The method described here is easy, fast and
detects, but does not locate, most leaks. It also verifies the sample pump condition.

1. Turn the analyzer ON, and allow at least 30 minutes for flows to stabilize.

2. Cap the sample inlet port (cap must be wrench-tight).

3. After several minutes, when the pressures have stabilized, note the SAMP (sample

pressure) and the RCEL (vacuum pressure) readings.