And no, Determination, And therefore no – Teledyne 9110T - Nitrogen Oxides Analyzer User Manual

Page 321: Which is defined here as the sum of no and no, Does not react with o

Advertising
background image

Model 9110T NOx Analyzer

Principles of Operation

Teledyne Analytical Instruments

321

Figure 13-1:

Reaction Cell with PMT Tube and Optical Filter

The narrowness of this band of sensitivity allows the 9110T to ignore extraneous light and radiation that
might interfere with the 9110T’s measurement. For instance, some oxides of sulfur can also be
chemiluminescent emitters when in contact with O

3

but give off light at much shorter wavelengths

(usually around 260nm to 480nm).

0

20

40

60

80

100

120

140

0.5

0.9

1.1

1.3

1.5

1.7

1.9

Wavelength

(micrometers)

In

tens

it

y

(ar

bi

tr

ar

y u

n

its

)

Optical Hi-Pass Filter Performance

NO + O

3

Emission Spectrum

Area of Sensitivity

Infrared 



Visible

Spectrum

PMT

Response

0.7

Figure 13-2:

9110T Sensitivity Spectrum

13.1.3. NO

X

AND NO

2

DETERMINATION

The only gas that is actually measured by the 9110T is NO. NO

2

, and therefore NO

x

(which is defined

here as the sum of NO and NO

2

in the sample gas), contained in the gas is not detected because NO

2

does not react with O

3

to create chemiluminescence.