88

Chapter 3 Administrative Tasks

4. Choose the

CalFactor item by pressing the adjacent soft key. The following items

appear on the generator’s display:

5. Choose the

Self-Calibrate item by pressing the adjacent soft key.

During calibration, calibration factors and other information appear on the generator’s
display. Observe the generator’s display for error messages. If an error message
appears, make a note of it and call your technical support representative for further
assistance.

To self-calibrate the generator using the command line interface:

1. Establish a session with the generator using either HyperTerminal over a serial

connection or Telnet over an Ethernet LAN. See

Establishing a terminal session with

the generator

” on page 31 or

Establishing a Telnet session with the generator

” on page

34.

2. At the prompt, enter the following command:

SCAL

The generator begins the calibration procedure.

To view analog calibration factors:

Enter the following command to view the analog video red, green, blue calibration factors:

CALF

?

The generator presents the following in response:

1089, 2589, 1993, 3223, 1148, 2633, 2059, 3279, 1000, 2505, 1990, 3219,
793, 787, 3382, 3509

For an explanation of the calibration factors, see the

CALF

? command in the Appendix A,

Command Reference

.”

Calibrating frequency

All frequencies associated with the generator are derived from a common frequency
reference, which can be calibrated using the rate calibration command

RATC

.

To calibrate all frequencies:

1. Load a standard format and set the horizontal rate to 100 kHz.

For example, enter the following command:

FMTL

DMT0660

Edit-Factors

Self-Calibrate

Advertising