Quantum Data 881 User Manual

Page 644

Advertising
background image

626

Chapter 18 Testing HDCP

CPTX:DUTT

3

// specifies the device under test as a repeater

for test 3AB.

You can query the current value of the parameter with:

CPTX:DUTT

?

// returns the device under test type.

11. (Optional) Specify the remaining parameters in the

EditPCP menu using the table

above.

Alternatively, you can specify the parameters through the command line as follows:

CPTX:RPFT

0

// indicates whether DUT (repeater) supports 1.1

features such as Advanced Cipher and Enhanced
Link Verification. 1 = yes; 0 = no.

CPTX:RKSV

10

// specifies maximum number of downstream devices

that can be supported in the repeater’s KSV list

CPTX:ROOR

0

// indicates whether DUT (repeater) outputs outputs

content to the downstream repeater that does not
have any downstream device connected.
1 = yes; 0 = no.

12. Press the

Options key to save the capabilities definition.

The generator LCD will display the message “Saved”.

13. Press the

Tools key get back to the Reports menu.

The generator LCD will display the message “Saved”.

14. Select

HDCP to access the HDCP compliance test menu.

The following is displayed on the generator’s LCD.

15. Select

!CompRpt to initiate the HDCP compliance test.

The message “HDCP Compliance Test” is shown and then all the tests are shown in
sequence.

Alternatively, you can run the tests using the following command:

CPTX:CPTR

46

// specifies that all applicable tests will be run.

CPTX:CPTU

// Initiates the execution of the test

Refer to

To view the HDCP compliance report:

” on page 651 for procedures on how to

view the generated report.

16. Monitor the 882E LCD (with release 2.17.x) for a set of prompts during the 3A-01 and

3A-05 tests.

!CompRpt

EditPCP