881/882 Video Test Generator User Guide

(Rev A.22)

609

13. Press the

Tools key get back to the Reports menu.

The generator LCD will display the reports menu as shown below.

14. Select

HDCP to access the HDCP compliance test menu.

The following is displayed on the generator’s LCD.

15. Select

!CompRpt to initiate the HDCP compliance test.

The message “HDCP Compliance Test” is shown and then all the tests are shown in
sequence.

Refer to

To view the HDCP compliance report:

” on page 651 for procedures on how to

view the generated report.

Alternatively, you can run the tests using the following command:

CPTX:CPTR 46

// specifies that all applicable tests will be run.

CPTX:CPTU

// Initiates the execution of the test

16. For the 882CA (0nly), run the 1B-01 test through the command line for both settings of

the SDCZ parameter which indicates whether the 882 should use a downstream device
count of 0 or the device count specified by the SKSV parameter (Source Max KSV).

a. Set the SDCZ parameter to indicate that the 882 should use a downstream device count

of 0:

CPTX:SDCZ 0

// indicates that the 882 should use a downstream

device count of 0.

b. Run the 1B-01 test using the following command:

CPTX:CPTR 10

// Selects specific test (e.g. 1B_01) see table below

CPTX:CPTU

// Initiates the execution of the test.

c. Set the SDCZ parameter to indicate that the 882 should use the device as specified by

the SKSV parameter (Source Max KSV):

CPTX:SDCZ 1

// indicates that the 882 should use the device count

as specified by the SKSV command (Source Max KSV).

d. Run the 1B-01 test using the following command:

CPTX:CPTR 10

// Selects specific test (e.g. 1B_01).

CPTX:CPTU

// Initiates the execution of the test.

EDID

Packets

Misc

HDCP

!CompRpt

EditPCP