Working with the network interface, Establishing a telnet session with the generator – Quantum Data 881 User Manual

Page 52

Advertising
background image

34

Chapter 1 Getting Started

To configure the generator’s serial port through the command line interface:

1. Establish a session with the generator using HyperTerminal over a serial connection or

Telnet over an Ethernet LAN. For instructions, see

Establishing a terminal session with

the generator

” on page 31 and

Establishing a Telnet session with the generator

” on

page 34.

2. At the session prompt, enter the following command to query the generator for the

current serial port settings:

MODE

?

The generator returns the current values:

9600,N,8,1,N,N

3. To change the settings, enter the following command:

MODE

baud parity data stop handshake protocol

For example, to change the baud rate to 38400, enter the following command:

MODE

38400 n 8 1 n n

Note: In this example, after you press Enter, the baud rate of the session and baud rate
of the generator will no longer match. This will cause the session to lose its connection
with the generator. Close the session, change the session baud rate to 38400, and then
re-open the session.

Working with the network interface

This section describes how to connect the generator to a PC via an Ethernet LAN and how
to establish a Telnet session with the generator.

To connect the generator to a PC over an Ethernet LAN:

1. Connect the Ethernet cable between the PC’s Ethernet port and an active Ethernet

jack.

2. Connect an Ethernet cable between the generator’s ETHERNET port and an active

Ethernet jack.

Establishing a Telnet session with the generator

The following procedure describes how to establish a Telnet session with the generator
over an Ethernet LAN. For information about establishing a terminal session over a serial
connection, see

Establishing a terminal session with the generator

” on page 31.

To establish a Telnet session with the generator:

1. Using a text terminal application, such as DOS Command Prompt, enter the following

command:

Advertising
This manual is related to the following products:

882