Quantum Data 881 User Manual

Page 653

Advertising
background image

881/882 Video Test Generator User Guide

(Rev A.22)

635

To run the 3B-01 HDCP compliance test on an HDMI repeater:

For the 882E releases prior to 2.17.x and for the 882CA, the 3B-01 tests are required
to be run with 2 different set ups through the command line for both read configurations:
1) short read and 2) long read.

1. For the 882E analyzer connect the cables as follows:

a. Apply power to the Encryption Status Tester (EST) using the power adapter

supplied with the EST.

b. Connect an HDMI cable from the repeater device under test output to the EST

HDMI port labeled, “HDMI IN.”

c. Connect an HDMI cable from a compliant source device output to the input of the

repeater device under test.

d. Connect an HDMI cable from the 882 HDMI Rx 1 port to the EST HDMI port labeled

“HDMI OUT.”

e. Connect an HDMI cable from the 882 HDMI Tx 2 port to the EST HDMI port labeled

“HDMI.”

f.

Connect an HDMI cable from the 882 HDMI Tx 1 port to the HDMI input of the
display device (used to monitor the video).

The following diagram depicts the test setup for the 882EA:

2. Ensure that the 882 settings used for the 3B-0x series described in

To run the 3B

series HDCP compliance tests on an HDMI repeater:

” on page 628, are maintained

27

3B_03 (Repeater Downstream
w/Repeater)

Irregular Procedure: (Second Part of
Authentication) Verify V’

28

3B_04 (Repeater Downstream
w/Repeater)

Irregular Procedure: (Second Part of
Authentication) MAX_DEVS_EXCEEDED

29

3B_05 (Repeater Downstream
w/Repeater)

Irregular Procedure: (Second Part of
Authentication) MAX_CASCADE_EXCEEDED

CPTX:CPTR

I

ndex

Test

Description

Advertising