Operation – Leica Biosystems ASP6025 User Manual

Page 100

Advertising
background image

98

Instructions for Use V 2.2 RevI – 08/2013

Tables with warning threshold values

The RMS system – adjusting warning threshold values

(continued)

The tables in this section list the recommended limits/warning threshold values for common
reagents approved for the ASP6025.
The warning threshold values recommended here must be used only in combination with the
preinstalled processing protocols. For other processing protocols created by the user, corresponding
threshold values have to be determined by the lab itself.

The warning threshold values in the following tables can be used as initial values for instrument
setup. These are to be used according to the preinstalled protocols by entering them into the

SET UP REAGENTS & WARNING VALUES menu.

5. Operation

These warning threshold values are set to 55 % at the factory. However, the user

can freely select between 50 % and 60 % as needed.

Autorotation protocols

Reagent

recommended replacement

Formalin

after 600 cassettes

Water

after 1200 cassettes

70 % ethanol

- threshold value 55 %

1 xylene

after 1500 cassettes

1 paraffin

after 1800 cassettes

Cleaning xylene

after 6 cycles

Cleaning alcohol

after 6 cycles

These warning threshold values must be validated before use, i.e. tissue pro-

cessing with patient tissue for diagnostics, by the laboratory itself according to

the local or regional accreditation requirements.