81

Leica ASP6025

The Service functions menu

(continued)

• The ASP6025 software continuously records which reagent was the last

reagent in the retort. The last reagent in the retort is displayed next to the
button. This way, it is ensured that incompatible reagents do not come into
contact with one another.

• If a filling process was canceled, it is possible that two reagents are displayed

here.

• If the user knows that there actually was a different reagent in the retort than

the one that is displayed, this can be corrected by selecting the LAST RETORT

REAGENT button.

• After the button has been pressed, first a warning message appears which

must be confirmed. Then, the current reagent list is displayed and it is thus
possible to change the reagent status of the retort by selecting the correct
reagent.

Reagent status of the retort

Xylene

The status of the PDF printer is displayed here. It cannot be changed.
"Printer ready" means that a PDF file can be generated and saved after a
program has run, if this was specified in the run options (see also

Chapter 5.1.2,

Fig. 94

).

Display of printer status

Fig. 107

Printer

ready

5. Operation

Fig. 108

Caution!

The reagent status of the retort may not be changed unless it

is known with certainty that the displayed reagent is not cor-

rect. If incompatible reagents come into contact with one an-

other, this can result in a malfunction of the instrument or the

impairment of a process.

Advertising