Cleaning and maintenance – Leica Biosystems ASP6025 User Manual

Page 161

Advertising
background image

159

Leica ASP6025

6. Cleaning and Maintenance

Replacing the active carbon filter

• The life of the active carbon filter will depend on the reagent

types used and how frequently they are used in the instru-
ment.

• The filter (13) should be replaced at least every 30 days

(

Fig. 227

).

After replacing the filter, open the SYSTEM MONITOR menu

to set the filter replacement date to the current date. See also

Chapter 5.1.4.

Fig. 227

Emptying the condensate bottle

• The blue condensate bottle must be taken out, emptied and

reinserted at regular intervals (weekly). This should be done
sooner if a prompt to that effect is issued by the software
(

Fig. 225

).

Fig. 225

Filter label

14

13

Fig. 226

Dispose of waste solvents with care according

to local regulations and the waste manage-

ment policy of the company or institution.

When inserting the new filter, make sure that

it is pushed all the way into the compartment

and that the flap is closed.