3 cleaning protocols, Cleaning and maintenance – Leica Biosystems ASP6025 User Manual
Page 155
153
Leica ASP6025
6. Cleaning and Maintenance
6.1.3 Cleaning protocols
I.
Rapid cleaning program
Step:
Reagent:
Time (min):
Temp. (°C):
Pressure/Vacuum:
1
Cleaning xylene
6
67
Ambient
2
Cleaning alcohol
3
62
Ambient
Dry step:
Empty retort
3
Vacuum step
10
67
Vacuum
4
Ventilation step
2
Ambient
Ambient
Optional step instead of "Dry step":
5
Cleaning water
1
65
Ambient
II.
Standard cleaning program
Step:
Reagent:
Time (min):
Temp. (°C):
Pressure/Vacuum:
1
Cleaning xylene
12
67
Ambient
2
Cleaning alcohol
6
62
Ambient
Dry step:
Empty retort
3
Vacuum step
10
67
Vacuum
4
Ventilation step
2
Ambient
Ambient
Optional step instead of "Dry step":
5
Cleaning water
1
65
Ambient
III.
Extended cleaning program
Step:
Reagent:
Time (min):
Temp. (°C):
Pressure/Vacuum:
1
Cleaning xylene
25
67
Ambient
2
Cleaning alcohol
10
62
Ambient
3
Cleaning agent
16
65
Ambient
4
Cleaning water
16
65
Ambient