79

Leica ASP6025

Fig. 104

Active carbon filter

The System monitor menu

(continued)

Also in Supervisor mode, only the runtime of the activated charcoal
filter can be reset in the SYSTEM MONITOR menu (

Figs. 103/104

).

If the ACTIVE CARBON FILTER button is touched, a field appears
which displays the programmed warning and alarm threshold values.
Even in supervisor mode, neither of these values can be changed.
• If the stored warning value is exceeded, a warning message

regarding the age of the carbon filter is output.

This warning serves as a reminder to replace the carbon filter. You

can continue using the instrument without any limitations.

• If the set alarm threshold value is exceeded, a message is output

on the screen to inform the user that the charcoal filter age has
been exceeded.

• When this message appears, at the latest, the active carbon filter

must be replaced in order to ensure that it functions and works
efficiently.