51

Leica ASP6025

4. Instrument Setup

• The drip tray (16) is located below the reagent cabinet. It

collects reagents that have overflowed or spilled so that no
pollution can form in or under the instrument. The volume
is approx. 5 l .

• The drip tray (16) must be regularly checked for signs of

reagents that have leaked. To do so, pull the tray out by the
handle (

Fig. 53

) and empty it if necessary.

Reagent cabinet with system bottles (

continued

)

4.4.9 The drip tray

16

Fig. 53

Condensate bottle
• The condensate bottle (14,

Fig. 52

) has the exact same shape

and connection design as the system bottle. The difference
is that it is made of blue plastic.

• It captures and collects condensate that forms in the system.

A mark for the maximum level is on the front side.

• Regularly check and empty the condensate bottle (once a

week) (see

Chapter 6.2

.2).

Fig. 52

14

Dispose of waste solvents with care according

to local regulations and the waste manage-

ment policy of the company or institution.