• Attach the exhaust flange (75) to the fan plate (78

in

Fig. 22

) and fasten it using the screws that held

the plate.

• First, screw in all four screws (77) loosely, then

tighten them diagonally using the size 3 Allen key
(76) to a tightening torque of 0.5 Nm (

Figs. 22, 23

).

Make sure that the fan plate and the flange are flush
against each other.

Installing the tube for external exhaust air

(continued)

Fig. 23

Fig. 24

• Now attach one end of the exhaust hose (74) to the

upwards-pointing tube (79,

Fig. 23

) of the flange

and push it down as far as it will go (

Fig. 24)

.

• Finally, connect the other end of the exhaust hose

with the external exhaust station.