Leica ASP6025
• Attach the exhaust flange (75) to the fan plate (78
in
Fig. 22
) and fasten it using the screws that held
the plate.
• First, screw in all four screws (77) loosely, then
tighten them diagonally using the size 3 Allen key
(76) to a tightening torque of 0.5 Nm (
Figs. 22, 23
).
Make sure that the fan plate and the flange are flush
against each other.
Installing the tube for external exhaust air
(continued)
Fig. 23
Fig. 24
• Now attach one end of the exhaust hose (74) to the
upwards-pointing tube (79,
Fig. 23
) of the flange
and push it down as far as it will go (
Fig. 24)
.
• Finally, connect the other end of the exhaust hose
with the external exhaust station.
75
76
79
77
77
74