161

Leica ASP6025

Checking the retort lid seal
• Regularly check the seal for damage after cleaning. If the seal is damaged, it must replaced

without delay.

6. Cleaning and Maintenance

Cleaning the retort sieve
• The filter sieve in the base of the retort (

Fig. 229

) must be cleaned with ethanol or xylene. For

convenience, the sieve may be taken out to remove all solid dirt.

• Call up CLEAN TOUCHSCREEN under PROGRAM

PROTOCOLS on the SETTINGS MENU screen.

• Clean the touchscreen.

• Press ENABLE EXIT to enable the BACK button (

Fig. 230

).

This will enable you to return to the standard display.

Fig. 230

Only the following cleaners can be used to clean the screen:
• Surface cleaner for plastic (Poliboy)
• Paraguard by Pseudoscience
• PC monitor cleaner (spray)

Cleaning the touchscreen

This way, you will be able to touch and clean

the screen without inadvertently enabling a

function.